Cited 0 times in Scipus Cited Count

Correction of target-controlled infusion following the wrong selection of emulsion concentration of propofol

DC Field Value Language
dc.contributor.advisor채, 윤정-
dc.contributor.author이, 원일-
dc.date.accessioned2015-10-28T23:56:59Z-
dc.date.available2015-10-28T23:56:59Z-
dc.date.issued2015-
dc.identifier.urihttp://repository.ajou.ac.kr/handle/201003/11841-
dc.description.abstractBackground: We investigated the correction methods following wrong-settings of emulsion

concentrations of propofol as a countermeasure against erroneous target-controlled infusions

(TCI).

Methods: TCIs were started with targeting 4.0 μg/ml of effect-site concentration (Ceff) of

propofol, and the emulsion concentrations were selected for 2.0% instead of 1.0% (FALSE1-

2, n = 24), or 1.0% instead of 2.0% (FALSE2-1, n = 24). These wrong TCIs were corrected at

3 min after infusion start. During FALSE1-2, the deficit was filled up while injecting after

equilibrium (n = 12), or while overriding (n = 12). During FALSE2-1, the overdose was

evacuated while targeting Ceff (n = 12) or targeting plasma concentration (Cp) (n = 12). The

gravimetrical measurements of TCI reproduced the Cp and Ceff using simulations. The

reproduced Ceff at 3 min (Ceff-3min) and the time to be normalized within ± 5% of target Ceff

(T±5%), were compared between the correction methods.

Results: During the wrong TCI, Ceff-3min was 1.98 ± 0.01 μg/ml in FALSE1-2, and 7.99 ± 0.05

μg/ml in FALSE2-1. In FALSE1-2, T±5% was significantly shorter when corrected while

overriding (3.9 ± 0.25 min), than corrected after equilibrium (6.9 ± 0.05 min) (P < 0.001). In

FALSE2-1, T±5% was significantly shorter during targeting Cp (3.6 ± 0.04 min) than targeting

Ceff (6.7 ± 0.15 min) (P < 0.001).

Conclusions: The correction methods, based on the pharmacokinetic and pharmacodynamic

characteristics, could effectively and rapidly normalize the wrong TCI following erroneously

selections of the emulsion concentration of propofol.
-
dc.description.tableofcontentsⅠ.INTRODUCTION 1

Ⅱ.MATERIALS AND METHODS 2

Ⅲ.RESULTS 7

Ⅳ.DISCUSSION 11

Ⅴ.CONCLUSION 15
-
dc.language.isoen-
dc.titleCorrection of target-controlled infusion following the wrong selection of emulsion concentration of propofol-
dc.title.alternative목표농도주입법에서 프로포폴 농도를 잘못 설정 후 교정 방법-
dc.typeThesis-
dc.identifier.urlhttp://dcoll.ajou.ac.kr:9080/dcollection/jsp/common/DcLoOrgPer.jsp?sItemId=000000019017-
dc.subject.keywordDrug delivery systems-
dc.subject.keywordInfusion pumps-
dc.subject.keywordIntravenous infusion-
dc.subject.keywordPropofol-
dc.subject.keyword약물 전달 시스템-
dc.subject.keyword주입 펌프-
dc.subject.keyword정맥 주입-
dc.description.degreeMaster-
dc.contributor.department대학원 의학과-
dc.contributor.affiliatedAuthor이, 원일-
dc.date.awarded2015-
dc.type.localTheses-
dc.citation.date2015-
dc.embargo.liftdate9999-12-31-
dc.embargo.terms9999-12-31-
Appears in Collections:
Theses > School of Medicine / Graduate School of Medicine > Master
Files in This Item:
There are no files associated with this item.

qrcode

해당 아이템을 이메일로 공유하기 원하시면 인증을 거치시기 바랍니다.

Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.

Browse