Comparison of Ocular Pulse Amplitude Measured Using Dynamic Contour Tonometry and Ocular Blood Flow Analyzer

Other Title
Dynamic Contour Tonometry와 안혈류분석계의 안구 박동 크기 비교
Authors
Kim, YH | Lee, MV | Ahn, JH
Citation
Journal of the Korean ophthalmological society, 56(12). : 1906-1912, 2015
Journal Title
Journal of the Korean ophthalmological society
ISSN
0378-64712092-9374
Abstract
Purpose: To compare ocular pulse amplitude (OPA) measured using dynamic contour tonometry (DCT) and ocular blood flow analyzer (BFA).

Methods: Thirty-five eyes of 35 patients were enrolled in this cross-sectional and retrospective study. OPA was measured using DCT. Pulse amplitude (PA) and pulsatile ocular blood flow were measured using BFA.

Results: OPA measured using DCT (2.79 ± 0.89 mm Hg) was not significantly different from PA measured with BFA (3.02 ± 0.90 mm Hg; p = 0.082) and both were significantly correlated (r = 0.663, p < 0.001). Mean difference ± limit of agreement was -0.22 ± 1.44 mm Hg between OPA and PA. OPA correlated significantly with intraocular pressure (IOP) measured using Goldmann applanation tonometry (r = 0.330, p = 0.047) but not PA (r = 0.057, p = 0.745). Both PA and OPA did not show significant correlation with the spherical equivalent of refractive error and central corneal thickness.

Conclusions: Although both OPA and PA measure IOP fluctuation and are not significantly different, they showed different relationships with IOP.
Keywords
DOI
10.3341/jkos.2015.56.12.1906
Appears in Collections:
Journal Papers > School of Medicine / Graduate School of Medicine > Ophthalmology
Ajou Authors
안, 재홍
Files in This Item:
10.3341_jkos.2015.56.12.1906.pdfDownload
Export

qrcode

해당 아이템을 이메일로 공유하기 원하시면 인증을 거치시기 바랍니다.

Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.

Browse