Cited 0 times in Scipus Cited Count

Clinical efficacy of custom-built software for the early detection of glaucoma: A comparison of axial-length and major retinal artery location data

Authors
Jang, H | Lee, SM | Ahn, J  | Rho, S
Citation
Korean journal of ophthalmology : KJO, 33(2). : 103-112, 2020
Journal Title
Korean journal of ophthalmology : KJO
ISSN
1011-89422092-9382
Abstract
Purpose: To assess the clinical efficacy for early detection of glaucoma using custom-built image software visualizing translucent retinal nerve fiber layer thickness (RNFLT) that is graphed based on a normative database. Methods: This prospective study was conducted using a normative database constructed with RNFLT data of 151 healthy Korean eyes. The reference lines of the mean, the lower 5%, and the lower 1% limit were visualized as a translucent RNFLT graph produced by our software after inputting each subject’s major retinal artery position and overlaying the results onto the RNFLT measurements. Fifty-eight additional healthy control and 79 early-glaucoma eyes were collected for the validation group. If a subject’s RNFLT graph was outside the reference line of the lower 1% limit, the graph was defined as abnormal. The lower 1% limit, which was generated by three criteria (criterion 1, built-in software; criterion 2, axial-length data; criterion 3, major retinal artery data), was used to address the difference of agreement with a standard answer. Results: For criteria 1, 2, and 3, the accuracy of our custom-built software was significantly higher than that of the manufacturer’s database (kappa of 0.475 vs. 0.852 vs. 0.940; sensitivity of 62.0% vs. 91.1% vs. 97.5%, respectively) maintaining high specificity (87.9% vs. 94.8% vs. 96.6%, respectively). Conclusions: The custom-built imaging software with the constructed RNFLT normative database showed high clinical efficiency for early detection of glaucoma with negligible user-related variability. © 2019 The Korean Ophthalmological Society
Keywords

DOI
10.3341/KJO.2018.0060
Appears in Collections:
Journal Papers > School of Medicine / Graduate School of Medicine > Ophthalmology
Ajou Authors
안, 재홍
Files in This Item:
10.3341_KJO.2018.0060.pdfDownload
Export

qrcode

해당 아이템을 이메일로 공유하기 원하시면 인증을 거치시기 바랍니다.

Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.

Browse