dc.description.abstractPURPOSE: To determine the effect of suction duration on corneal flap thickness and incision angle of the cut margin created by a microkeratome in a porcine eye.



METHODS: Thirty porcine eyes were randomly assigned to three groups according to different suction duration: group 1 (10 sec), group 2 (35 sec), and group 3 (60 sec). The Hansatome microkeratome with a #160 plate and 8.5-mm-diameter suction ring was used to create a corneal flap with a diameter of 8.5 millimeters. Corneal flap thickness was measured by automated ultrasonic pachymetry, and the incision angle was assessed by measuring the angle of a sagittal section of the cornea using image analysis software.



RESULTS: Mean corneal flap thickness in groups 1 (10 sec), 2 (35 sec), and 3 (60 sec) was 87.8 +/- 22.0 microm, 116.0 +/- 7.0 microm, and 127.2 +/- 16.8 microm, respectively. There was a statistically significant difference between groups 1 (10 sec) and 2 (35 sec) (P=.005) and groups 1 and 3 (P=.004). The mean incision angle in groups 1, 2, and 3 was 34.8 +/- 9.0 degrees, 44.4 +/- 16.1 degrees, and 48.24 +/- 15.3 degrees, respectively. A statistically significant difference was found between groups 1 (10 sec) and 2 (35 sec) (P=.044) and groups 1 (10 sec) and 3 (60 sec) (P<.001).



CONCLUSION: In laser in situ keratomileusis in porcine eyes, an increase in suction duration resulted in a thicker flap and greater incision angle.
-
dc.language.isoen-
dc.subject.MESHAnimals-
dc.subject.MESHCornea-
dc.subject.MESHCorneal Topography-
dc.subject.MESHKeratomileusis, Laser In Situ-
dc.subject.MESHMicroscopy, Video-
dc.subject.MESHOphthalmologic Surgical Procedures-
dc.subject.MESHSuction-
dc.subject.MESHSurgical Flaps-
dc.subject.MESHSwine-
dc.subject.MESHTime Factors-
dc.titleEffect of microkeratome suction duration on corneal flap thickness and incision angle.-
dc.typeArticle-
dc.identifier.pmid12458865-
dc.contributor.affiliatedAuthor장, 재우-
dc.type.localJournal Papers-
dc.citation.titleJournal of refractive surgery (Thorofare, N.J. : 1995)-
dc.citation.volume18-
dc.citation.number6-
dc.citation.date2002-
dc.citation.startPage715-
dc.citation.endPage719-
dc.identifier.bibliographicCitationJournal of refractive surgery (Thorofare, N.J. : 1995), 18(6). : 715-719, 2002-
dc.relation.journalidJ01081597X-
Appears in Collections:
Journal Papers > School of Medicine / Graduate School of Medicine > Ophthalmology
Files in This Item:
There are no files associated with this item.

qrcode

해당 아이템을 이메일로 공유하기 원하시면 인증을 거치시기 바랍니다.

Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.

Browse