Cited 0 times in Scipus Cited Count

Stability of Specimens for N-mid Osteocalcin and intact Osteocalcin Assays and Correlation of Results between the Two Assays

DC Field Value Language
dc.contributor.author정, 민권-
dc.contributor.author임, 영애-
dc.contributor.author곽, 연식-
dc.date.accessioned2012-02-13T05:12:11Z-
dc.date.available2012-02-13T05:12:11Z-
dc.date.issued2000-
dc.identifier.issn1225-097X-
dc.identifier.urihttp://repository.ajou.ac.kr/handle/201003/5621-
dc.description.abstractBackground : Serum osteocalcin has been known to be the marker of bone metabolism. NMIDTM Osteocalcin One Step ELISA (BioTech A/S, Denmark, for N-mid OC) measures intact osteocalcin (1-49 a.a.) including N-terminal midfragments (1-43 a.a.) whereas NovoCalcin (Metra Biosystems, USA, for intact OC) measures intact osteocalcin. We evaluated the stability of specimens and the correlation both assays.
Method : We used the sera of 189 women over 40 years old for the correlation of N-mid OC and intact OC. The following specimens were prepared to evaluate the stability of specimens: whole blood and serum specimens were allowed to stand at ambient temperature for 0, 3 and 6 hours. Pooled sera from 3 apparently healthy individuals were stored at 41C, -201C and -701C and were then subjected to be assayed in 0, 1, 2, 4 weeks after the storage. The assays were done in triplicate in a single analytical batch.
Results : The means±S.D. (ng/ml) of N-mid OC and intact OC were 9.79±5.76 and 6.43± 3.71, respectively. The correlation of both assays were as follows, r=0.79 (P<0.005), and [intact OC] = 0.51 x [N-mid OC] + 1.45. The storage of whole blood samples at room temperature showed gradual decrease of results from both assays, i.e. 91 % and 89% of baseline value 3 and 6 hours after for N-mid OC and 79% and 75% for intact OC. However, the serum samples were stable at room temperature for the same duration; 107% and 98% for N-mid OC and 94% and 103% for intact OC, respectively. The results of 1, 2 and 4 weeks after storage of serum specimens at 41C showed a significant decrease; 76%, 62% and 54% for N-mid OC and 21%, 9% and 0% for intact OC; those at -20´C showed 110%, 108% and 77% for N-mid OC and 111%, 85% and 105% for intact OC; and those at -701C showed 108%, 107% and 92% for N-mid OC and 108%, 95% and 110% for intact OC.
Conclusions : The results of serum N-mid OC correlate well with that of intact OC. The serum specimens were stable at room temperature for 6 hours, but whole blood samples were unstable. Therefore, serum should be separated without delay after phlebotomy. Serum should be frozen at - 701C for storage longer than 1 week.
-
dc.formatapplication/pdf-
dc.language.isoko-
dc.titleStability of Specimens for N-mid Osteocalcin and intact Osteocalcin Assays and Correlation of Results between the Two Assays-
dc.title.alternative혈청 N-mid osteocalcin과 intact osteocalcin의 상관성 및 검체 안정성 조사-
dc.typeArticle-
dc.identifier.urlhttp://kmbase.medric.or.kr/Main.aspx?d=KMBASE&m=VIEW&i=0375320000220020243-
dc.subject.keywordN-mid osteocalcin-
dc.subject.keywordIntact osteocalcin-
dc.subject.keywordmarker of bone metabolism-
dc.subject.keywordstability-
dc.subject.keywordstorage-
dc.contributor.affiliatedAuthor정, 민권-
dc.contributor.affiliatedAuthor임, 영애-
dc.type.localJournal Papers-
dc.citation.titleJournal of clinical pathology and quality control-
dc.citation.volume22-
dc.citation.number2-
dc.citation.date2000-
dc.citation.startPage243-
dc.citation.endPage250-
dc.identifier.bibliographicCitationJournal of clinical pathology and quality control, 22(2). : 243-250, 2000-
dc.relation.journalidJ01225097X-
Appears in Collections:
Journal Papers > School of Medicine / Graduate School of Medicine > Laboratory Medicine
Files in This Item:
Journal of clinical pathology and quality control_22(2)_243-250.pdfDownload

qrcode

해당 아이템을 이메일로 공유하기 원하시면 인증을 거치시기 바랍니다.

Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.