Cited 0 times in Scipus Cited Count

Optical immunosensor for quantifying C-telopeptide fragments of type II collagen as an osteoarthritis biomarker in urine.

Authors
Kim, SJ | Park, YM | Min, BH  | Lee, DS | Yoon, HC
Citation
Biochip journal, 7(4). : 399-407, 2013
Journal Title
Biochip journal
ISSN
1976-02802092-7843
Abstract
A fluorescence-based biosensor for measuring the quantity of C-telopeptide fragments of type II collagen (CTX-II) was developed as an osteoarthritis biomarker in urine. During osteoarthritis progression, joint components such as cartilage collagen are degraded by collagenase-protease and secreted into the serum as CTX-II. To detect CTX-II in urine (uCTX-II) having structural heterogeneity, conventional sandwich format assay does not provide sufficient accuracy. As an alternative, a competitive immunoassay was developed for the quantification of uCTX-II, which uses antibody-conjugated fluoro-microbeads to generate an optical signal. After preparing the biosensing surface, the uCTX-II sample and antibody-optical probe conjugates were reacted on the PEG4-EKGPDP exposed biosensing surface. The optical probes competed with the uCTX-II in the sample for binding to the immobilized PEG4-EKGPDP. This assay was able to detect uCTX-II concentrations between 200 ng/mmol (corrected value vs. creatinine) and 1,400 ng/mmol, encompassing the clinical detection ranges required for osteoarthritis diagnosis. The competitive immunoassay developed for uCTX-II detection was rapid and exhibited good correlation with conventional ELISA methods. This novel competition assay is a promising tool for the diagnosis and monitoring of osteoarthritis using urine samples.
Appears in Collections:
Journal Papers > School of Medicine / Graduate School of Medicine > Orthopedic Surgery
Ajou Authors
민, 병현
Files in This Item:
There are no files associated with this item.
Export

qrcode

해당 아이템을 이메일로 공유하기 원하시면 인증을 거치시기 바랍니다.

Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.

Browse